GC-4000A Series Gas Chromatographs can be used in analysis of a wide range of chemicals, from ambient temperature gases to liquids and gas-liquid mixtures with boiling points of 400℃ or lower. This extremely versatile instrument has many applications in industries such as oil, coal, chemical engineering, environmental protection, textile, agriculture, medicinal research, and sanitation.
Technical specifications:
|
TCD |
FID |
ECD |
FPD |
NPD | |
|
Sensitivity |
S≥5000 |
|
|
|
|
|
Detectability |
|
M≤1×10-11 |
M ≤1×10-13 |
Mp≤2×10-12 |
MN≤5×10-13 |
|
Noise |
≤0.1mv |
≤5×0-14A |
≤0.01mv |
≤2×10-12A |
≤2×10-13A |
|
Align range |
|
106 |
103 |
103 |
103 |
|
Excursion |
|
1×10-14 |
0.02mv/30min |
3×10-13 |
2.5×10-12 A/h |
1. We offer five types of detectors. Each is designed to fit the customers' special needs.
√ Flame Ionization Detector (FID): used for highly sensitive organic compound analysis.
√ Thermal Conductivity Detector (TCD): analyzes ambient temperature gases, water, and small hydrocarbon compounds.
√ Flame Photometric Detector (FPD): used in analysis of pesticide traces, sulfur and phosphorous compounds.
√ Electron Capture Detector (ECD): for detecting pesticides, and chlorine-containing compounds.
√ Nitrogen-Phosphorus Detector (NPD): ideal for analyzing trace quantities of nitrogen and phosphorus.
2. With unit type monolithic construction design, it is easy to the installation and maintenance.
3. Outfitted with large column oven, three columns can be installed simultaneously.
4. Reasonably designed, the instrument may install three kinds of detectors of different types simultaneously. In addition, it has three independent electro circuit systems and three signal output fan-outs.
5. Flexible packed column and capillary column system.
6. Stable temperature control system; 8 ramp programmable heating; the sample injector and every detector may control the temperature independently and has over temperature protection function.
7. Fire new regulating and multi-gas system; single, double and three gas path can be chosen discretionarily.
Injector and valve sample injection system are outfitted or chosen discretionarily.
Reversible flow and pre incision device can be chosen discretionarily.
Methane reforming furnace and cracking furnace are to be chosen freely.
Whole set of instruments
• Instrument (model can be chosen freely);
• Data treatment workstation: PC, printer, special software and interface can be chosen.
• Air supply: Air generator, hydrogen generator and high purity nitrogen gas can be chosen.
• Chromatographic column is to be chosen in line with the requirement of the clients.
• Spare parts.
The followings can be chosen for chromatographs.
GH-200(G-101) High purity hydrogen generator
GH-100(G-101A) High purity hydrogen generator
GA-100(G-103) High purity air generator
GC-4000A Series Data Sheet (Please click here)
GC-4000A Gas Chromatographs (Big Screen Liquid Crystal Display)
Technical characteristics
1. Column oven
Volume of column oven: 300×300×200mm
Temperature ranges: indoor temperature +5℃~400℃
Temperature veracity: 1% of the enactment value.
Overheat protection: discretional temperature within 400℃
Precision of the temperature control: ±0.1℃
2. Big screen liquid crystal display will display temperature, exterior event, operational conditions, etc.
3. Automatic back door control; temperature exceeding protection.
4. Constant temperature enactment of man-machine conversation, or 8 ramps programmable heatingup.
Temperature ranges: indoor temperature +5℃~400℃
Velocity of heatingup: 0.1~40℃/min.
Precision of the temperature control: ±0.1℃
Windage of temperature indication value: 1%
5. Heatingup program
Programmable ramp: 8 ramps
Velocity of heatingup: 0~40℃/min (increment: 0.1℃/min)
6. Way of sample injection
Capillary diffluence/sample injection of non diffluence, packed column sample
injection, cold chapiter sample injection.